IVF Sensor Modules
The Okolab IVF sensor modules, combined with LEO hand held meter, are user-friendly meters specifically designed to monitor the critical parameters of the IVF laboratory.
The sensor modules are calibrated by using traceable standards, and measure with high accuracy in all working range. Calibration data are stored in their local memory. The sensor modules can be delivered with traceable or accredited calibration.
The sensor modules can be sent to Okolab for re-calibration. Moreover calibration can be performed by activating an intuitive user calibration routines.
Each sensor can perform single point measurement, logging, and store data in LEO’s local memory. The stored data can be displayed on LEO’s screen, organized by date and device. The stored data can be downloaded through the micro USB port and analyzed by using the LEO DATA IMPORT Excel Macro.
T1-MODULE
Small, flexible, calibrated, and immersible thermocouple. Ideal for measurements in liquids, such as culture media in a dish. T1-MODULE has a negligible diameter (0.13mm) and it can be directly introduced into an incubator, even if it lacks a dedicated temperature measurement port.
Features
- Sensor: K-type thermocouple
- Measurement range: 0-60°C
- Accuracy: ±0.1°C from 20 to 50°C
- Working temperature: 0-60°C
- 5-Point calibration performed by using traceable calibration standards
T2-MODULE
Calibrated temperature probe, ideal for measurements of incubator temperature when sampling port access is available. T2-MODULE has an external diameter of 2 mm and is suitable for temperature measurements of benchtop and high-top incubators where the temperature monitoring port has a diameter larger than 2 mm.
Features
- Sensor: PT1000 RTD Class A
- Measurement range: 0-60°C
- Accuracy: ±0.1°C from 20 to 50°C
- Working temperature: 0-60°C
- 5-Point calibration performed by using traceable calibration standards
CO2-PPM-MODULE
Designed to measure carbon dioxide at ppm in ofces, laboratories, hospitals, and generally, wherever indoor air quality is important for the occupants’ comfort and health. Traceable calibration standards are used in order to ensure high accuracy in the range 0-2000 ppm. The probe is temperature and pressure compensated and can be used to adjust any fxed CO2 measurement system.
Features
- Sensor: Non-dispersive infrared (NDIR)
- Measurement range: 0-10000 ppm
- Accuracy: ±(50ppm +3% reading)
- Working temperature: 0-50°C
- Response time: 2 minutes
- 4-Point calibration performed by using traceable calibration standards
HUMIDITY-MODULE
Designed for checking measurements in humidity chambers, incubators, laboratories, and to monitor indoor air quality. RH-MODULE can be used as a reference to adjust any humidity instruments.
Features - Relative Humidity
- Sensor:Hygromer HT-1
- Measurement range: 0-100%
- Accuracy at 23°C: ±0.8% RH
- Working temperature: -10 +100°C
- Response time: 18 s
- 3-Point calibration performed by using traceable calibration standards
Features - Temperature
- Sensor: Pt100 1/3DIN Class B
- Measurement range: -10 + 100°C
- Accuracy: ±0.2°C from 20 to 40°C
- Working range: -10 +100°C
- 1-Point calibration performed by using traceable calibration standards
FLOW-RATE-MODULE
Compact and accurate flow meter with short response time, ideal for instantaneus gas flow measurements from benchtop incubator chambers. The highly compact and lightweight design makes it ideal for many different work environments. The presence of display allows to use the sensor module also as a standalone device.
Features
- Sensor:MEMS
- Measurement range: 0-200 ml/min
- Accuracy: ± (2.0 reading +0.5% full scale)
- Repeatability: ±0.5% full scale
- Response time:
- Working pressure: 0.08-0.8 bar
- Working temperature: -10 + 55°C
- Calibration medium: air
- 10-Points calibration performed by using traceable calibration standards