H601-NIKON-TI-SR-GLASS/FRAME
H601-NIKON-TI-SR-GLASS/FRAME is a dual channel plate featuring Indium Tin Oxide (ITO) coated glass cased in a heated aluminium frame.
It provides a warm and optically transparent surface perfectly flush with the microscope stage with the best possible temperature uniformity and accuracy, achieved by controlling the temperature of both the glass and the metal frame.
H601-NIKON-TI-SR-GLASS/FRAME features anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass cased in rigid aluminum frame to ensures the highest resistance against breakage and scratches.
Protected by 10-year warranty on glass damage.
It requires H401-T-CONTROLLER and uses both control channels (one for the glass and one for the flame).
Features
- Flush with Nikon T/Ti-SR and Ti-SP stage
- Dual channel: heated Glass and heated Metal Frame
- Temperature range: from ambient to 60°C
- Accuracy on specimen: ±0.3°C
- ITO glass coating ensures uniform glass heating
- Glass thickness: 0.5 mm
- Aluminum frame improves rigidity
- Anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass
- 10-year warranty on glass damage
Temperature Uniformity
assessed via Infra Red Imaging
Heated Glass - Heated Frame
Superior Temperature uniformity and stability obtained by heating both the glass and the metal frame
H401-T-CONTROLLER
2-Channel Temperature Controller with 4.3'' touch screen interface compatible with all H401 heating devices. One controller can control independently up to two items.
External temperature sensor (included) and self calibration routines allows easy calibration of the heating devices for the highest accuracy on sample temperature.
On board data logging and download procedures to USB drives allow to record temperature as a function of time and to log calibration events and alarms.
The controller is equipped with password protected routines to re-calibrate the sensors against certified thermometers.
Features
- T - range: from ambient to 60°C
- T - accuracy on specimen: ± 0.3°C
- 4.3'' touch screen interface
- External T - sensor (included)
- On-board Data logging and download routines to USB drive
- On-board logging of calibration and alarm events
- Alarm buzzer and External Alarm Connector
- One controllers can control independently up to two devices
- Self calibration routines allow to improve temperature accuracy
- Password protected procedure for periodic calibration against certified thermometer
Touch Pad Interface
Beautifully designed, simple to use and very compact.
T - range: from ambient to 60°C
4.3'' touch screen interface
Beautifully designed, simple to use and very compact.
External T - sensor (included)
On-board Data logging and download routines to USB drive
One controllers can control independently up to two devices
Self calibration routines allow to improve temperature accuracy
Take advantage of the dual channel option
One channel controls the metal T-Pad.
The other channel controls the Glass Plate.
Data Logging
On-Board Data Logging - stored in local memory
A simple procedure allows to download data to USB.